Application
Bacterial Retention
Quantitative Retention of 1 × E7 CFU/cm²
Bacterial Endotoxins
< 0.25 EU/mL as determined by the LAL test
Extractables
The total amount of extractables is well below the limits established by the current USP unter "Sterile Water for Injection".
Connector Inlet
1/2" Single Stepped Hose Barb
Connector Outlet
Filter Size
size 8 (0.1 m² / 1.08 ft²)
Filtration Area
0.1 m² / 1.08 ft²
Outer Diameter
108.5 mm (77 mm without Connection / Valves)
Capsule/Cartridge Format
Cartopore
Maximum Allowable Back Pressure
2 bar (29 psi) at 20°C; 1.5 bar (21.8 psi) at 50°C
Maximum Diffusion
4 mL/min at 2.5 bar | 36 psi
Minimum Required Bubble Point
3.2 bar | 46 psi
Vent Valve
Standard
Filter Material
Cellulose Acetate (CA)
Housing Material
Polypropylene (PP)
Material o-ring | seal
Silicone (SI)
Support Fleece
Maximum Differential Pressure
20 °C: 5 bar; 50 °C: 3 bar
Maximum Operating Pressure
5 bar (72.5 psi)
Non-Fiber Releasing
This filter product complies with the title 21 of the Code of Federal Regulations (CFR), section 210.3(b)(6) and 211.72.
Pore Size Final-filter
0.2 µm
Pore Size Pre-filter
0.45 µm
Brand
Sterilization Method
Autoclavable
Sterilization Procedures
Autoclaving: at 134°C, 30 min, 2 bar, min. 25 Cycles